Etelux Glove Box Regeneration Techniques

From
Jump to: navigation, search

1) Regeneration gasoline: a mixture of hydrogen and inert gasoline (hydrogen accounts for five% -ten%)

two) Amount: 40L fuel cylinder 10Mpa

3) Regenerative fuel tension: .05Mpa (the fuel consumption for every 50 % hour in the state of .05Mpa is that the cylinder pressure drops by 1.5Mpa. If it exceeds one.5Mpa, the force of the pressure regulator’s program will be decreased, if it is less than 1.5Mpa Boost the strain of the stress regulator program)

four) Method: 3 hours heating-three hrs heating and ventilation-three hours vacuuming-1 hour pumping-10 hrs cooling

five) Measures:

① Join the regeneration gasoline (examine the pipeline for leaks after link: immediately after connecting, open the key valve of the strain decreasing valve to the utmost, open the auxiliary meter to .1Mpa, transform off the main valve, wait 15 minutes, if the strain of the auxiliary meter continues to be at .one Mpa continues to be unchanged, indicating no air leakage

>②Set the stress of the auxiliary gauge of the strain reducing valve to .1Mpa (the pressure during ventilation is about .06-.08MPa

>③ Close the cycle and analyz

r>④Turn on the vacuum glovebox pu

r>⑤ Get started regenerat

br>⑥ The new application finishes at 20h (clean right before cleaning, drinking water and oxygen index <200p


6) Note: Do not cut off the power and gas during the regeneration process (especially the first 6 hours)! Some pressure reducing valves are not allowed, and the pressure needs to be adjusted appropriately to ensure continuous ventilation for 3 ho


7) Observation: whether the regeneration gas flowmeter shows about 20L / min, otherwise adjust the pressure of the pressure reducing valve.